WF 3169 Battery Stack Monitor Module

WF 3169 Battery Stack Monitor Module

Is the sample rate specified for full configuration?
Answer: Yes, it is.  The specified rate of 235 Hz means that all 24 channels will be sampled 235 times per second. So you will get 24 measurement values ev...
Thu, 9 Jul, 2020 at 11:47 AM
Are all measurements taken simultaneously?
Answer: No, there is is a set of four multiplexers and ADC’s in the hardware that scans the inputs in groups of four with an offset of six channels. Fir...
Thu, 9 Jul, 2020 at 12:55 PM
What should I do if I have fewer battery cells than channels on the WF3168 or WF3169 module?
If you don't have enough battery cells to fill all the channels of the module you will have to put jumpers (a short length of conductor used to close ci...
Wed, 9 Jun, 2021 at 4:54 PM